syringe_pumps_ne

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

syringe_pumps_ne [2023/01/30 20:54] – created bdincausyringe_pumps_ne [2023/01/30 21:42] (current) bdincau
Line 3: Line 3:
 ^                     Syringe Pumps                             ^ ^                     Syringe Pumps                             ^
 |          {{ :tools:syringepumps:syringe_pumps.jpg?nolink&200 |}}      | |          {{ :tools:syringepumps:syringe_pumps.jpg?nolink&200 |}}      |
-| **Tool Type:**                  Microfluidics                           +| **Tool Type:**                  Single Syringe Pump                          
-| **Location:**                   Microfluidics Lab                           |+| **Location:**                   Microfluidics Lab - Elings 3430                          |
  
  
-| **Description:**                Syringe pump                           |+| **Description:**                Single Syringe Pump for droplet generation and other milli/microfluidic experiments                           |
 | **Manufacturer:**               New Era                           | | **Manufacturer:**               New Era                           |
  
Line 14: Line 14:
 ===== About ===== ===== About =====
  
 +Syringe pumps utilize a stepper motor, lead screw, and pusher block to dispense fluid from a syringe at a controlled rate. These New Era pumps can accommodate only one syringe at a time. If you need dual syringe capabilities, please check out the [[syringe_pumps_ha| Harvard Apparatus Syringe Pumps]]
  
 +Syringe pumps are flow-rate-controlled devices, which means that you program the pump to operate at a fixed flow rate regardless of the pressure required. (As opposed to pressure-controlled devices which fix the pressure.)
 ---- ----
  
Line 20: Line 22:
 ===== Detailed Specifications ===== ===== Detailed Specifications =====
  
 +See product manual for extended specifications.
  
 +Maximum force: 35 lb (155 N) \\ 
 +Minimum syringe size & minimum flow rate: 1 mL @ 0.73 uL/hr \\ 
 +Maximum syringe size & maximum flow rate: 60 mL @ 1.7 L/hr \\ 
  
 ---- ----
  
 +===== Safety Concerns =====
 +
 +**Pinching Hazard** - Be careful when loading syringes to avoid pinching oneself with the syringe holders. \\ 
 +**High Pressure Hazard - Always wear goggles when operating syringe pumps.** If too much pressure accumulates in your system, often due to clogging, your tubing or attached fluidic devices can rupture. This can create a jet of fluid which presents a physical hazard to those nearby. \\ 
 +
 +----
 +
 +===== Operating Procedures =====
 +
 +Please reference the attach operation manual for detailed operating instructions.
 +
 +**General Operation** \\ 
 +1) Select your syringe size using either the measured internal diameter of the syringe. \\ 
 +2) Install and fill your syringe. \\ 
 +3) Select the desired flow rate. \\
 +4) Select the desired target volume (optional). If you set a target volume and operate in volume mode, pumping will automatically cease when the target volume has been reached. Otherwise, the pump will not stop until manually turned off. \\ 
 +5) When finished, remove syringes, power off the pump, and clean the pump if contaminated by any fluid during your experiment.
 +
 +----
  