===== About ===== | ===== About ===== | ||
+ | Syringe pumps utilize a stepper motor, lead screw, and pusher block to dispense fluid from a syringe at a controlled rate. These New Era pumps can accommodate only one syringe at a time. If you need dual syringe capabilities, | ||
+ | Syringe pumps are flow-rate-controlled devices, which means that you program the pump to operate at a fixed flow rate regardless of the pressure required. (As opposed to pressure-controlled devices which fix the pressure.) | ||
---- | ---- | ||
Line 20: | Line 22: | ||
===== Detailed Specifications ===== | ===== Detailed Specifications ===== | ||
+ | See product manual for extended specifications. | ||
+ | Maximum force: 35 lb (155 N) \\ | ||
+ | Minimum syringe size & minimum flow rate: 1 mL @ 0.73 uL/hr \\ | ||
+ | Maximum syringe size & maximum flow rate: 60 mL @ 1.7 L/hr \\ | ||
---- | ---- | ||
+ | ===== Safety Concerns ===== | ||
+ | |||
+ | **Pinching Hazard** - Be careful when loading syringes to avoid pinching oneself with the syringe holders. \\ | ||
+ | **High Pressure Hazard - Always wear goggles when operating syringe pumps.** If too much pressure accumulates in your system, often due to clogging, your tubing or attached fluidic devices can rupture. This can create a jet of fluid which presents a physical hazard to those nearby. \\ | ||
+ | |||
+ | ---- | ||
+ | |||
+ | ===== Operating Procedures ===== | ||
+ | |||
+ | Please reference the attach operation manual for detailed operating instructions. | ||
+ | |||
+ | **General Operation** \\ | ||
+ | 1) Select your syringe size using either the measured internal diameter of the syringe. \\ | ||
+ | 2) Install and fill your syringe. \\ | ||
+ | 3) Select the desired flow rate. \\ | ||
+ | 4) Select the desired target volume (optional). If you set a target volume and operate in volume mode, pumping will automatically cease when the target volume has been reached. Otherwise, the pump will not stop until manually turned off. \\ | ||
+ | 5) When finished, remove syringes, power off the pump, and clean the pump if contaminated by any fluid during your experiment. | ||
+ | |||
+ | ---- | ||