- | ==== Workshop Wizard | + | ===== RETIRED |
- | Microfluidics | + | ====Elings 3430 Lab Maintenance |
__Chores:__ | __Chores:__ | ||
Line 8: | Line 8: | ||
1. //Check and replenish paper towels in dispensers// | 1. //Check and replenish paper towels in dispensers// | ||
* Paper towels are stocked in brown paper packages on the shelves along the wall by the entrance. | * Paper towels are stocked in brown paper packages on the shelves along the wall by the entrance. | ||
- | * There are 2 paper towel dispensers in the lab, just turn the key to check how much there are. | + | * There are 2 paper towel dispensers in the lab uFL and 3 in IW, just turn the key (usually located on top of the dispenser) |
2. //Check and replenish gloves// | 2. //Check and replenish gloves// | ||
- | * Check if any glove boxes are about to run out, should be 2 glove locations. | + | * Check if any glove boxes are about to run out; there are 2 glove locations |
- | * If they are, open up the next one on top of the stack and add a new box of gloves to the bottom of the stack. | + | * If they are, open up the next one on top of the stack and add a new box of gloves to the bottom of the stack. Insure there are always 2 extra boxes in the main glove location located by the lab entrance. |
- | * Make sure there are both nitrile and latex gloves to use. | + | * Make sure there are both latex gloves to use. |
- | * Gloves are stocked on the shelves along the wall by the entrance, as well as above the vacuum furnace. | + | * Gloves are stocked on the shelves along the wall by the entrance, as well as above the vacuum furnace |
3. //Tidy up bench spaces// | 3. //Tidy up bench spaces// | ||
* Properly dispose of sharps and put away tools/ | * Properly dispose of sharps and put away tools/ | ||
+ | * Ensure razor blades are located on designated magnets and __not__ on lab benches. | ||
4. //Sweep and mop// | 4. //Sweep and mop// | ||
Line 24: | Line 25: | ||
* Dirty water poured in the sink | * Dirty water poured in the sink | ||
* Use the general maintenance cart located in the hallway outside the lab. The cleaner is stored in the cabinet under the sink. | * Use the general maintenance cart located in the hallway outside the lab. The cleaner is stored in the cabinet under the sink. | ||
- | * **Rule of thumb:** Best to do in the mornings or weekends. | + | * **Rule of thumb:** Best to do in the mornings or weekends |
- | 5. //Check and replenish hazardous waste labels// | + | 5. //Check and replenish hazardous |
+ | - Empty solvent bottles should __not__ be rinsed. Make sure to cross out the labels with a Sharpie and have a waste solvent label pre tagged. | ||
+ | - If waste bin is HALF full, [[https:// | ||
+ | * **Note**: The hazardous waste tub is the grey tub located underneath the sink. | ||
6. //Check sharps containers// | 6. //Check sharps containers// | ||
Line 34: | Line 38: | ||
7. //Check glass waste bin// | 7. //Check glass waste bin// | ||
* If full, seal and tape off the top. | * If full, seal and tape off the top. | ||
+ | * Replace with new bin. | ||
+ | * To take out the glass waste, there are rolling trash bins with red tops behind Elings Hall. Ensure that two people take out the glass waste, and that you hold the bins from the bottom. | ||
- | 8. //Make sure all cabinets are locked// | + | 8. //Make sure all cabinets are locked |
+ | * If a user has left it unlocked, make sure to lock it. | ||
9. //Check and replenish Kim wipes at every station// | 9. //Check and replenish Kim wipes at every station// | ||
Line 41: | Line 48: | ||
10. //Refill solvent squirt bottles in fume hood// | 10. //Refill solvent squirt bottles in fume hood// | ||
- | * Fume hood should have 3 bottles of IPA, 2 bottles of Ethanol, Acetone. | + | * Fume hood should have 1-3 bottles of IPA and Acetone, and 1-2 bottles of Ethanol. |
- | * Large jugs under the hood in the cabinet. | + | * Large jugs should be located |
* New jugs in the cabinet labeled "extra solvents." | * New jugs in the cabinet labeled "extra solvents." | ||
* Wear gloves and lab coat; do all refilling in the fume hood. | * Wear gloves and lab coat; do all refilling in the fume hood. | ||
Line 50: | Line 57: | ||
11. //Monitor the hazardous waste jugs// | 11. //Monitor the hazardous waste jugs// | ||
* Place in the hazardous waste bin if needed. | * Place in the hazardous waste bin if needed. | ||
+ | * Check waste levels in the containers. | ||
+ | * Generally, move to pickup bin if the hazardous waste jug has not been replaced for about 9 months __or__ if it is over 75% full. | ||
12. //Top off the Sonicator with DI water// | 12. //Top off the Sonicator with DI water// | ||
Line 76: | Line 85: | ||
17. //Clean the pressure washing station// | 17. //Clean the pressure washing station// | ||
* Make sure water is off (both levers should be down). | * Make sure water is off (both levers should be down). | ||
- | * Check the yellow gloves and replace torn. | + | * Check the yellow gloves and replace |
18. //Restock Thinky Mix Uline jars// | 18. //Restock Thinky Mix Uline jars// | ||
- | * Stored below the workbench. | + | * Stored below the workbench |
19. //Replace dirty bench liners// | 19. //Replace dirty bench liners// | ||
Line 92: | Line 101: | ||
22. //Refill CNC drill squirt bottle// | 22. //Refill CNC drill squirt bottle// | ||
* Use blue coolant. | * Use blue coolant. | ||
+ | * Check the level in the jug and refill if necessary. | ||
+ | * Undiluted coolant is underneath the sink. Mix the cool according to the ratio written on the bottle. | ||
23. //Clean the laser cutter// | 23. //Clean the laser cutter// | ||
Line 108: | Line 119: | ||
- | **Innovation Workshop | + | ====Elings 2442 Maintenance |
- | ** | + | |
- | I) Check gloves: extra boxes are in microfluidics lab, write down what is taken on clipboard on the fume hood nearest to the exit. | + | 1. //Sweep, mop, and vacuum the lab// |
+ | *The broom and dustpan | ||
+ | *The mop is located upstairs in uFl and the cleaning fluid is under the sink near the door. | ||
+ | *The shop-vac should be emptied weekly and the batteries replaced as needed | ||
- | 2) Check paper towels/WypAll: there is a dispenser | + | 2. //Check and replenish |
+ | *There are paper towel dispensers | ||
+ | *Extra paper towels can be found in uFl. | ||
- | 3) Clean up sander/grinder: Use paintbrush hanging with the tools and the vacuum to clean up debris. | + | 3. //Check and replenish gloves// |
+ | *There are three locations to check for gloves: by the front door, by the IPA cleaning station, | ||
+ | *Extra boxes are located in uFl. | ||
- | 4) Lower fume hood sashes, except one with imaging equipment | + | 4. //Replenish hazardous waste labels// |
+ | *Empty solvent bottles should not be rinsed. Make sure to cross out the labels | ||
+ | *If the waste bin is half full, contact EH&S to schedule a waste pickup. | ||
- | 5) Put away tools | + | 5. //Check the sharps bins and glass waste bins// |
+ | *If full, seal and tape off the top and replace with a new bin. | ||
+ | *To take out the glass waste, there are rolling trash bins with red tops behind Elings Hall. Ensure that two people take out the glass waste, and that you hold the bins from the bottom. | ||
- | 6) Return lab coats and safety glasses | + | 6. //Take out the trash// |
+ | *The trash should be enclosed properly (wraps found in the drawer to the left of the hood) and marked as contaminate with liquid resin. | ||
+ | *Replace the container frequently to leave room for wrapping the trash. | ||
- | 7) Organize scrap wood: put it on shelf near lab coats | + | 7. //Tidy up and organize spaces// |
+ | *Put away tools, lab coats, and safety glasses. | ||
+ | *Make sure all drawers are closed and lower fume hood sashes (except the one with imaging equipment). | ||
+ | *Organize scrap wood: put it on the shelf near lab coats. | ||
+ | *When cleaning the electronics workbench, clear any wires and tools from the bench and check to make sure that the soldering ions are turned off. | ||
- | 8) Sweep and mop lab: Mop is upstairs, cleaning fluid is under sink near the door, broom is in the hallway in between labs | + | 8. //Make sure all power tools and cabinets are locked// |
- | 9) Replenish hazardous waste labels | + | 9. //Fill out the IPA in the resin cleaner to the halfway point of the indicator lines. Measure the specific gravity of the IPA in the cleaner// |
+ | *If the O-ring on the measuring device sits above the large flaps, replace the IPA. | ||
+ | *Reminder: Check the IPA level in the squirt bottles regularly. | ||
- | 10) Check sharps bins, also check to make sure there are no sharp metal scraps in the trash bins | + | 10. //Clean up sander/ |
+ | *Use the paintbrush hanging with the tools and the vacuum to clean up debris. | ||
- | 11) Check glass waste | + | 11. //Clean and lubricate the mill and the lathe// |
+ | *Vacuum away any chips and debris. | ||
+ | *Oil using the oil bottle into all lubrication valves. | ||
- | 12) Make sure all drawers are closed | + | 12. //Clean the bandsaw and the drill press// |
+ | *Use the vacuum. | ||
+ | *Ensure that you take the power supply off the drill press. | ||
+ | *On the bandsaw, open up the top door and make sure the inside is clean. | ||
- | 13) Make sure all power tools and cabinets are locked | + | 13. //Camera maintenance// |
+ | *Change the camera battery. The battery door is on the bottom of the camera | ||
+ | *Turn off the camera box light. The power button is on the side. | ||
- | 14) Clean and lubricate | + | 14. //Clean the laser cutter// |
+ | *The platform should be cleaned weekly to clear remnants of jobs. (Open the drawer for vacuuming if necessary) | ||
+ | *The lens should be cleaned weekly. | ||
+ | *Both of the mirrors (there are 2 mirrors) should be cleaned monthly. The cleaning log is found on the exhaust manager to the left of the cutter. | ||
- | 15) Clean the bandsaw | + | 15. // |
+ | *Make sure to wipe remaining glue on the Ultimaker platform. | ||
+ | *Take out the platform | ||
- | 16) Charge | + | 16. //Clean around |
+ | *Vacuum | ||
- | 17) Calibrate the scanner: | + | 17. //Formlabs maintenance// |
- | I) Go into the Einscan software | + | |
- | II) Choose calibrate | + | |
- | III) Place the Calibration stand on the turn-wheel and orient the calibration board (using the reference dots) on the stand as shown in the image on the software. | + | |
- | IV) Press snap. Once complete, re-orient | + | |
- | Note: The board and the stand can be found in the scanner fume hood. | + | |
- | 18) Turn off camera box light: the power button is on the side | + | |
- | 19) Thermo oven labels should be printed | + | 18. //Stratasys maintenance// |
+ | *Check the filament level, replace it with stock from the shelves in the corridor if necessary. | ||
+ | *To vacuum the print platform, go to settings | ||
- | 20) When cleaning | + | 19. //Miicraft maintenance// |
+ | *Remove | ||
+ | *Return the cover to protect from dust. | ||
+ | * **Note**: Be careful when stirring, Teflon is extremely sensitive. | ||
- | 21) If the build platform is not on the formlabs printer, use the additional rotation platform from the resin cleaning fume hood. | + | 20. //SCA 1200HT maintenance// |
- | + | *Check for the fluid threshold level weekly. Over longer intervals, remove | |
- | 22) The laser cutter cleaning has three components: | + | * **Note**: This device is particularly dangerous. Wear gloves, coat, and face shield when opening |
- | a) The platform should be cleaned weekly to clear remnants of jobs. (Open the drawer | + | * **Note**: |
- | b) The lens should be cleaned weekly. | + | |
- | c) The mirrors (2) should be cleaned monthly. | + | |
- | Cleaning log is found on the exhaust manager to the left of the cutter. | + | |
- | + | ||
- | 23) When clearing the cricut and ultimaker shelf, make sure to wipe remaining glue on the ultimaker platform. Take out the platform and use IPA. | ||
- | 24) The chips made by the carbide 3D should be regularly vacuumed. | + | ====Incubator chores==== |
- | Reminder: the shop vac should be emptied, and the batteries should be replaced on a weekly basis. | + | - Check hand soap dispenser - add soap if needed |
- | + | | |
- | 25) When replacing the liners in the 3D printer cleaning fume hood, use the stock in the cabinet located to the lower left of the hood. | + | |
- | + | ||
- | 26) Fill out the IPA in the resin cleaner to the halfway point of the indicator lines. | + | |
- | Measure the specific gravity of the IPA in the cleaner. If the O-ring on the measuring device sits above the large flaps, replace the IPA. | + | |
- | Reminder: | + | |
- | + | ||
- | 27) The trash should be enclosed properly (wraps found in the drawer to the left of the hood) and marked as contaminated with liquid resin. Replace the container frequently to leave room for wrapping the trash. | + | |
- | + | ||
- | 28) Stratasys maintenance: | + | |
- | a) Check the filament level, replace it with stock from the shelves in the corridor | + | |
- | b) To vacuum the print platform, go to settings and bring the gantry up; preferably use gloves. | + | |
- | + | ||
- | 29) Miicraft maintenance: | + | |
- | Note: Be careful when stirring, Teflon is extremely sensitive. | + | |
- | + | ||
- | 30) SCA 1200HT maintenance: | + | |
- | Note: This device is particularly dangerous. Wear gloves, coat, and face shield when opening the door. Ask Daniel or Mr. Bothman for detailed instructions. | + | |
- | Note: The alarm may sound if the container is lowered too quickly. Silence the alarm with the button that is indicated with a warning light but do so only if you are certain that this was the cause of the alarm. | + | |