User Tools

Site Tools


neytech_qex_furnace

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
neytech_qex_furnace [2023/01/05 21:19] – [Neytech Qex Furnace] haleyneytech_qex_furnace [2023/01/31 00:13] (current) – [Reference Documentation] bdincau
Line 3: Line 3:
  
 ^                     Neytech Furnace                               ^ ^                     Neytech Furnace                               ^
-|  {{ :tools:neytechqexfurnace:neytech_qex_furnace.jpg?nolink&200 |}}   | +|  {{ :Tools:neytech_furnace.jpeg?nolink&200 |}}   | 
-| **Tool Type:**                  Furnace                           |+| **Tool Type:**                  Vacuum Furnace                           |
 | **Location:**                   Microfluidics Lab                           | | **Location:**                   Microfluidics Lab                           |
  
  
  
-| **Description:**                "TEXT HERE"                           |+| **Description:**                Small programmable furnace (1200 C) with vacuum optional         |
 | **Manufacturer:**               Neytech                           | | **Manufacturer:**               Neytech                           |
  
Line 16: Line 16:
 ===== About ===== ===== About =====
  
-Insert Text Here! +This is an automatic furnace, ideal for material annealing. Six different programs can be created, each with their own temperature and gas flow. The device is normally stored in the Microfluidics Lab and must be moved to a bench top for operation. Contact the lab manager if you would like to use it.
- +
----- +
- +
-===== Detailed Specifications ===== +
- +
-Insert Text Here! +
 ---- ----
  
 ===== Safety Concerns ===== ===== Safety Concerns =====
 +**High Heat Hazard** - Never operate furnace near combustibles. Always wear thermally insulated gloves when transferring samples out of the furnace. \\ 
  
-Insert Text Here!+**Pinching Hazard** - Be careful when closing the top assembly, known as the muffle. It's possible to pinch your fingers/hands when closing. It is also possible to damage you sample if any part of your sample hangs off of the interior platform. \\ 
  
 ---- ----
  
 ===== Operating Procedures ===== ===== Operating Procedures =====
 +See attached operating manual for detailed operation. \\ 
  
-Insert Text Here!+General Operation \\  
 +1) Set up the furnace on an appropriate counter top. Make any desired vacuum or nitrogen connections.\\  
 +2) Plug the furnace in and power on.\\  
 +3) Set up your desired program.\\  
 +4) Open the muffle by pressing the double up arrow.\\  
 +5) Load your sample onto the platform, ensuring that none of you sample hangs over the edge.\\  
 +6) Close the muffle and run your program.\\  
 +7) When finished, move the muffle up, remove your sample, and power down the furnace.\\ 
  
 ---- ----
  
-===== Reference Documentation =====+===== Detailed Specifications ===== 
 +See attached manual for complete specifications\\  
 +\\  
 +Chamber Platform Diameter: 4 in (10 cm)\\  
 +Chamber Height: 2.5 in (6.3 cm)\\  
 +Max Temp: 1200 C\\ 
  
-Insert Text Here!+-----
  
-===== Training Documentation ===== 
  
-Insert Text Here! +===== Reference Documentation ===== 
-----+{{:Tools:neytech_man.pdf | Operation Manual}}
  
 +-----