User Tools

Site Tools


fisher_acumet

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
fisher_acumet [2023/01/05 23:04] – [Operating Procedures] haleyfisher_acumet [2024/09/26 20:17] (current) epace
Line 5: Line 5:
 |      {{ :tools:fisheracumet:fisher_acumet.jpg?nolink&200 |}}          | |      {{ :tools:fisheracumet:fisher_acumet.jpg?nolink&200 |}}          |
 | **Tool Type:**                  Measurement                           | | **Tool Type:**                  Measurement                           |
-| **Location:**                   Microfluidics Lab                          |+| **Location:**                   Elings 3430                          |
  
  
 | **Description:**                pH meter                           | | **Description:**                pH meter                           |
 | **Manufacturer:**               Fisher Scientific                          | | **Manufacturer:**               Fisher Scientific                          |
 +
 +1/5/23 Haley
  
 ---- ----
Line 16: Line 18:
 This pH meter allows you to measure pH, absolute mV, relative mV or temperature, select one of three sets of standard buffer groups, and standardize with up to 5/6 buffers. This pH meter allows you to measure pH, absolute mV, relative mV or temperature, select one of three sets of standard buffer groups, and standardize with up to 5/6 buffers.
  
 +To use this tool, please ask one of our staff for help (located in a locked cabinet beneath the Bambu Printer). The buffers are located in Chemical Storage Cabinet #4.
 ---- ----
  
Line 25: Line 28:
 conditioned and used in accordance with manufacturer’s conditioned and used in accordance with manufacturer’s
 instructions." instructions."
 +
 +"Because electrodes vary in their response, you must standardize or
 +calibrate your pH meter and electrode to compensate for electrode
 +variation. The more frequently you standardize or calibrate, the more
 +accurate your measurements. Standardize at least daily, or more
 +often, for accurate results."
 ---- ----
  
Line 47: Line 56:
   - Repeat steps 3-6 with the pH 10 (blue) buffer solution. When the meter accepts the second , pH 10, buffer solution, it will briefly display the percent slope associated with the electrode’s performance prior to returning to the Measure mode. If the electrode is within the range of 90 – 102%, the GOOD ELECTRODE message will appear.    - Repeat steps 3-6 with the pH 10 (blue) buffer solution. When the meter accepts the second , pH 10, buffer solution, it will briefly display the percent slope associated with the electrode’s performance prior to returning to the Measure mode. If the electrode is within the range of 90 – 102%, the GOOD ELECTRODE message will appear. 
  
 +__Measurement__
  
 +  - Immerse electrode into solution and stir moderately.
 +  - When the meter senses that the reading has stabilized, the stable icon will appear under the reading. This message will not appear if the stability indicator has been turned off during setup.
 +  - You can save your reading to the meter’s memory using the save key.
  
  