Line 53: Line 53:
   - Repeat steps 3-6 with the pH 10 (blue) buffer solution. When the meter accepts the second , pH 10, buffer solution, it will briefly display the percent slope associated with the electrode’s performance prior to returning to the Measure mode. If the electrode is within the range of 90 – 102%, the GOOD ELECTRODE message will appear.    - Repeat steps 3-6 with the pH 10 (blue) buffer solution. When the meter accepts the second , pH 10, buffer solution, it will briefly display the percent slope associated with the electrode’s performance prior to returning to the Measure mode. If the electrode is within the range of 90 – 102%, the GOOD ELECTRODE message will appear. 
  
 +__Measurement__
  
 +  - Immerse electrode into solution and stir moderately.
 +  - When the meter senses that the reading has stabilized, the stable icon will appear under the reading. This message will not appear if the stability indicator has been turned off during setup.
 +  - You can save your reading to the meter’s memory using the save key.
  
  