User Tools

Site Tools


process_info_pdms_devices

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
process_info_pdms_devices [2021/07/15 21:39] – [Device Bonding] magnusonprocess_info_pdms_devices [2024/09/25 22:26] (current) kho
Line 218: Line 218:
  
 **Procedure:** **Procedure:**
-  - Confirm that the oven is set to 80 C. Either oven in the Microfluidics Lab is acceptable. +  - Confirm that the oven is set to 80 C. Either oven in Elings 3430 is acceptable. 
   - Identify an empty oven tray to use for spill protection.   - Identify an empty oven tray to use for spill protection.
   - Carefully transport degassed elastomer to the ovens.   - Carefully transport degassed elastomer to the ovens.
Line 422: Line 422:
   - Remove the metal lid of the sonicator bath. Check that the bath is filled to the operating level (if not, fill to the line with DI).   - Remove the metal lid of the sonicator bath. Check that the bath is filled to the operating level (if not, fill to the line with DI).
   - Place the plastic cutouts across the bath, resting it on the ledges that seat the lid. One cutout has 2 openings, the other has only 1.   - Place the plastic cutouts across the bath, resting it on the ledges that seat the lid. One cutout has 2 openings, the other has only 1.
-  - Retrieve the appropriate beakers from the drying rack above the sink in the microfluidics lab.   - Select the beakers labeled Acetone, IPA, and DI Water. Confirm that each beaker has a plastic tube / zip tie ring around it to act as a cushion.+  - Retrieve the appropriate beakers from the drying rack above the sink in Elings 3430.   - Select the beakers labeled Acetone, IPA, and DI Water. Confirm that each beaker has a plastic tube / zip tie ring around it to act as a cushion.
   - Fill each beaker with the corresponding chemical using the solvent bottles in the cabinet underneath the fume hood.   - Fill each beaker with the corresponding chemical using the solvent bottles in the cabinet underneath the fume hood.
   - Carefully seat each beaker in one of the cutouts. Confirm that the base of the container rests below the water level.   - Carefully seat each beaker in one of the cutouts. Confirm that the base of the container rests below the water level.
process_info_pdms_devices.1626385154.txt.gz · Last modified: 2021/07/15 21:39 by magnuson