Previous revision
plasma_cleaner [2023/01/05 03:02] – [Harrick Plasma Cleaner PDC-32G] haleyplasma_cleaner [2024/11/07 22:02] (current) epace
Line 1: Line 1:
-====== Harrick Plasma Cleaner PDC-32G======+====== Harrick Plasma Chamber PDC-32G======
  
-^                           Plasma Cleaner                           ^+^                           Plasma Chamber                           ^
 |          {{:tools:plasmacleaner:1876642.jpg?200|}}                                                            |          {{:tools:plasmacleaner:1876642.jpg?200|}}                                                           
-| **Tool Type:**                  Cleaning Tool                   +| **Tool Type:**                  Surface treatment                  
-| **Location:**                   Microfluidics Lab                   |+| **Location:**                   Elings 3430                   |
  
  
-| **Description:**                Plasma Cleaner                           |+| **Description:**                Plasma chamber                           |
 | **Manufacturer:**               Harrick                           | | **Manufacturer:**               Harrick                           |
  
Line 16: Line 16:
  
 The Harrick PDC-32G is a compact table top plasma cleaner for nanoscale surface cleaning and activation of small samples. The attached PlasmaFlo gas mixer allows for quantitative control of up to two different process gasses and monitoring of the vacuumed pressure within the chamber. The Harrick PDC-32G is a compact table top plasma cleaner for nanoscale surface cleaning and activation of small samples. The attached PlasmaFlo gas mixer allows for quantitative control of up to two different process gasses and monitoring of the vacuumed pressure within the chamber.
 +
 +This tool is plumbed for nitrogen plasma only, which still works very well for PDMS-glass bonding and PDMS-PDMS bonding. If you require oxygen plasma, please consider using the [[https://www.cnsi.ucsb.edu/facilities/equipment/yes-r3-plasma-cleaning-system | plasma chamber in the Nanostructures Cleanroom Facility]] instead.
  
 ---- ----
  
 ===== Safety Concerns ===== ===== Safety Concerns =====
 +
 +The plasma chamber uses ** RF radiation**. If you have a pacemaker or similar device, please consult your doctor before using.
 +
 +
  
 ---- ----
Line 25: Line 31:
 ===== Training Documentation ===== ===== Training Documentation =====
  
-[[Plasma Cleaner SOP]]+{{ :tools:plasmacleaner:ufl_harrick_plasma_operating_instructions.pdf | Plasma Cleaner SOP}}
  
 ---- ----
  
 ===== Recipes ===== ===== Recipes =====
-Brian Dincau found that the following recipe does a good job enhancing PDMS-PDMS bonding:+Brian Dincau found that the following recipe works well for bonding PDMS to PDMS or Glass:
   * Power - High   * Power - High
-  * Pressure - 450 mTorr+  * Pressure - 440 to 460 mTorr
   * Duration - 30s   * Duration - 30s
   * Nitrogen flow - adjust as needed to maintain desired pressure.    * Nitrogen flow - adjust as needed to maintain desired pressure. 
 +  * Mate surfaces immediately after treatment, then transfer to an oven at 115C for 20 minutes.
  
 ---- ----