| ====== Fisher Acumet AB15 pH Meter ====== | ====== Fisher Acumet AB15 pH Meter ====== | ||
| - | The following is a sample image that can be inserted by clicking on the small picture frame icon in the edit bar. | ||
| - | ^ Fisher Acumet AB15 pH Meter | + | ^ pH Meter |
| - | | | | + | | {{ : |
| - | | **Tool Type: | + | | **Tool Type: |
| - | | **Location: | + | | **Location: |
| - | | **Supervisor** | **Tool Lead** | + | | **Description: |
| - | | David Bothman | + | | **Manufacturer: |
| - | | (805) 893-4125 | + | |
| - | | bothman@cnsi.ucsb.edu | + | |
| - | | **Description: | + | 1/5/23 Haley |
| - | | **Manufacturer: | + | |
| ---- | ---- | ||
| ===== About ===== | ===== About ===== | ||
| + | This pH meter allows you to measure pH, absolute mV, relative mV or temperature, | ||
| - | Insert Text Here! | + | To use this tool, please ask one of our staff for help (located in a locked cabinet beneath the Bambu Printer). The buffers are located in Chemical Storage Cabinet #4. |
| + | ---- | ||
| + | |||
| + | ===== Safety Concerns ===== | ||
| + | " | ||
| + | measurements. Improper care of the electrode may cause the | ||
| + | meter reading to drift, respond slowly, or produce erroneous | ||
| + | readings. For this reason, the electrode should always be | ||
| + | conditioned and used in accordance with manufacturer’s | ||
| + | instructions." | ||
| + | " | ||
| + | calibrate your pH meter and electrode to compensate for electrode | ||
| + | variation. The more frequently you standardize or calibrate, the more | ||
| + | accurate your measurements. Standardize at least daily, or more | ||
| + | often, for accurate results." | ||
| ---- | ---- | ||
| - | ===== Detailed Specifications | + | ===== Operating Procedures |
| - | Insert Text Here! | + | __Setup__ |
| + | - Connect electrode arm to the base | ||
| + | - Connect power cable to outlet | ||
| + | - Remove protective cover from electrode | ||
| + | - Before first use of a glass electrode, soak it for 2-4 hours in an electrode storage solution, pH 4 buffer, or KCI solution | ||
| + | - Remove the shortening cap from input 1 to connect the combination pH electrode. If a combination electrode isn’t used, connect the indicating pH electrode. Plug the reference electrode into ref 1. Also, install the ATC probe into the ATC 1. | ||
| + | - Rinse and blot-dry (don’t wipe) electrodes between each measurement using DI water. | ||
| + | - Between measurements, | ||
| - | ---- | + | __Standardization__ |
| - | ===== Safety Concerns ===== | + | - Press and release the mode key until your digital display indicates pH mode. This key toggles between the pH, mV and Rel mV modes. |
| + | - Press the setup key twice and then press the enter key to clear an existing standardization. | ||
| + | - Rinse the electrode with distilled water using a squeeze bottle and immerse the rinsed electrode into pH 4 (pink) buffer solution. | ||
| + | - Press std again to access the Standardization mode. The selected buffer group is displayed briefly. Wait for the reading to stabilize. | ||
| + | - Press std again to initiate standardization. The meter will automatically recognize the buffer (4.00 not 7.00) and then return to the Measure screen. | ||
| + | - Repeat steps 3-6 with the pH 10 (blue) buffer solution. When the meter accepts the second , pH 10, buffer solution, it will briefly display the percent slope associated with the electrode’s performance prior to returning to the Measure mode. If the electrode is within the range of 90 – 102%, the GOOD ELECTRODE message will appear. | ||
| - | Insert Text Here! | + | __Measurement__ |
| - | ---- | + | |
| + | | ||
| + | | ||
| - | ===== Operating Procedures ===== | ||
| - | Insert Text Here! | + | ===== Detailed Specifications ===== |
| + | |||
| + | Relative accuracy: ±0.01 | ||
| + | Range: -1.99 to 19.99 | ||
| + | |||
| + | ----- | ||
| - | ---- | ||
| ===== Reference Documentation ===== | ===== Reference Documentation ===== | ||
| - | Insert Text Here! | + | {{: |