fisher_acumet
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revisionNext revisionBoth sides next revision | ||
fisher_acumet [2023/01/05 22:42] – haley | fisher_acumet [2023/01/05 22:59] – [About] haley | ||
---|---|---|---|
Line 5: | Line 5: | ||
| {{ : | | {{ : | ||
| **Tool Type: | | **Tool Type: | ||
- | | **Location: | + | | **Location: |
Line 14: | Line 14: | ||
===== About ===== | ===== About ===== | ||
+ | This pH meter allows you to measure pH, absolute mV, relative mV or temperature, | ||
---- | ---- | ||
===== Safety Concerns ===== | ===== Safety Concerns ===== | ||
- | + | " | |
+ | measurements. Improper care of the electrode may cause the | ||
+ | meter reading to drift, respond slowly, or produce erroneous | ||
+ | readings. For this reason, the electrode should always be | ||
+ | conditioned and used in accordance with manufacturer’s | ||
+ | instructions." | ||
---- | ---- | ||
===== Operating Procedures ===== | ===== Operating Procedures ===== | ||
+ | - Connect electrode arm to the base | ||
+ | - Connect power cable to outlet | ||
+ | - Remove protective cover from electrode | ||
+ | - Before first use of a glass electrode, soak it for 2-4 hours in an electrode storage solution, pH 4 buffer, or KCI solution | ||
+ | - Remove the shortening cap from input 1 to connect the combination pH electrode. If a combination electrode isn’t used, connect the indicating pH electrode. Plug the reference electrode into ref 1. Also, install the ATC probe into the ATC 1. | ||
+ | - Rinse and blot-dry (don’t wipe) electrodes between each measurement using DI water. | ||
+ | - Between measurements, | ||
- | ---- | ||
===== Detailed Specifications ===== | ===== Detailed Specifications ===== | ||
+ | Relative accuracy: ±0.01 | ||
+ | Range: -1.99 to 19.99 | ||
----- | ----- | ||
+ | |||
fisher_acumet.txt · Last modified: 2023/01/05 23:09 by haley