-^             Fisher Acumet AB15 pH Meter                               ^+^             pH Meter                               ^
 |      {{ :tools:fisheracumet:fisher_acumet.jpg?nolink&200 |}}          | |      {{ :tools:fisheracumet:fisher_acumet.jpg?nolink&200 |}}          |
-| **Tool Type:**                  "TEXT HERE"                           | +| **Tool Type:**                  Measurement                           | 
-| **Location:**                   "TEXT HERE"                           |+| **Location:**                   Microfluidics Lab                          |
  
  
-|  **Supervisor**        |  **Tool Lead**  |      +| **Description:**                pH meter                           | 
-|  David Bothman          "WW Name" +| **Manufacturer:**               Fisher Scientific                          |
-|  (805) 893-4125        |  (###) ###-#### +
-|  bothman@cnsi.ucsb.edu  |  "WW Email" +
- +
-| **Description:**                "TEXT HERE"                           | +
-| **Manufacturer:**               "TEXT HERE"                           |+
  
 ---- ----
  
 ===== About ===== ===== About =====
 +This pH meter allows you to measure pH, absolute mV, relative mV or temperature, select one of three sets of standard buffer groups, and standardize with up to 5/6 buffers.
  
-Insert Text Here!+----
  
 +===== Safety Concerns =====
 +"Proper electrode care is fundamental to obtaining reliable pH
 +measurements. Improper care of the electrode may cause the
 +meter reading to drift, respond slowly, or produce erroneous
 +readings. For this reason, the electrode should always be
 +conditioned and used in accordance with manufacturer’s
 +instructions."
 ---- ----
  
-===== Detailed Specifications =====+===== Operating Procedures ===== 
 +  - Connect electrode arm to the base 
 +  - Connect power cable to outlet 
 +  - Remove protective cover from electrode  
 +  - Before first use of a glass electrode, soak it for 2-4 hours in an electrode storage solution, pH 4 buffer, or KCI solution 
 +  - Remove the shortening cap from input 1 to connect the combination pH electrode. If a combination electrode isn’t used, connect the indicating pH electrode. Plug the reference electrode into ref 1. Also, install the ATC probe into the ATC 1.  
 +  - Rinse and blot-dry (don’t wipe) electrodes between each measurement using  DI water. 
 +  - Between measurements, store conventional pH electrodes in storage solution, pH 4 buffer, or KCI solution. Always leave the filling hole of liquid filled and combination electrodes open. Refill when the level of solution gets below the manufacturer’s recommended level.
  
-Insert Text Here! 
  
----- 
  
-===== Safety Concerns =====+===== Detailed Specifications =====
  
-Insert Text Here!+Relative accuracy: ±0.01  
 +Range: -1.99 to 19.99
  
-----+-----
  