====== FormCure UV Curing Station ====== ^ FormCure ^ | {{ :tools:formcureuv:formcureuv.png?nolink&200 |}} | | **Tool Type:** Post-processing | | **Location:** Innovation Workshop | | **Description:** FormLabs UV curing chamber | | **Manufacturer:** FormLabs | Last updated 1/5/23 Haley ---- ===== About ===== After your FormLabs print has gone through the FormWash, the resin should be cured in the FormCure, a UV curing chamber. Each resin has a specific hardening curve that can be found within the reference material. This will help you indicate the ideal time and temperature to be set on the FormCure. ---- ===== Safety Concerns ===== * Allow at least half an hour for IPA to evaporate before placing part in FormCure * Allow time for the Form Cure to cool off before removing parts * Do not operate form cure with lid open ---- ===== Operating Procedures ===== - Any part to be cured needs to be washed in the FormWash with IPA BEFORE being placed in the form cure. Supports can be removed before or after curing process. - Wait 30 minutes after washing to allow all remaining IPA to evaporate - Check the reference material for selected resin to determine ideal time and temperature for curing - Use the dial on the front of the FormCure to set time and temperature - Place part within the FormCure-> press start. ----- ===== Detailed Specifications ===== * Max cure temp: 80 degrees C * Light: 39W total drawn by 13 405 nm LEDs ---- ===== Reference Documentation ===== [[form2sop|Form 2 Training SOP]] [[https://support.formlabs.com/s/article/Using-Form-Cure?language=en_US]] [[https://support.formlabs.com/s/article/Form-Cure-Time-and-Temperature-Settings?language=en_US]] [[https://support.formlabs.com/s/article/Form-Cure-Time-and-Temperature-Settings?language=en_US]]